    Diagnostic problems in two-dimensional shear wave elastography of the liver

    2020-06-15 06:29:18HirokoNaganumaHideakiIshidaAtsushiUnoHiroshiNagaiHidekatsuKurodaMasahiroOgawa
    World Journal of Radiology 2020年5期

    Hiroko Naganuma,Hideaki Ishida,Atsushi Uno,Hiroshi Nagai,Hidekatsu Kuroda,Masahiro Ogawa

    Abstract

    Key words: Two-dimensional shear wave elastography;Ultrasound;Artifacts;Liver cirrhosis;Liver tumor;Computer simulation model

    INTRODUCTION

    Although liver biopsy has long been the standard method for diagnosing chronic liver disease,it is associated with many problems,such as postbiopsy complications[1],sampling error[2],and interobserver variability[3]. In a positive aspect,these problems motivated the development of noninvasive techniques to assess liver fibrosis,including magnetic resonance elastography[4-6]and ultrasound (US)-based elastography[7-10]. The latter has attained preferential use because of its lower cost and easier manipulation of the instrument[7-10]. Currently,there are four kinds of US-based elastography,namely strain elastography[11],transient elastography[12],point shear wave elastography[13],and two-dimensional shear wave elastography (2D-SWE)[9,14,15].Among these methods,2D-SWE has emerged as the most frequently used diagnostic US tool for hepatic fibrosis quantification;this is due to its ability to sample a large area in the liver,changing the sampling area quickly under B-mode observation and displaying a color mapping of shear wave (SW) values over the B-mode image[9,15].Measurement of SW values can be easily performed but it is also well known that the obtained SW values change according to the instrument itself and the level of expertise of the operator;as such,the certitude of those SW results is not always satisfactory[6,9,15]. The other reasons for this instability have not been clarified.

    This review reexamines the variability,using a method that is different from those in the commonly reported studies. US artifacts are routinely encountered in clinical practice and include refraction artifacts[16-18],attenuation artifacts[16,19,20],reflection artifact[16,21],reverberation artifacts[22]and others[16,23,24]. As use of 2D-SWE becomes more widespread,the problems related to US artifacts in it will be more important[25-27]. The aim of this review was,thus,to present a wide spectrum of US artifacts in hepatic 2D-SWE and to provide the simplest way to minimize the effects of such.

    THEORETICAL BASIS OF TWO-DIMENSIONAL SHEAR WAVE ELASTOGRAPHY

    The push-pulse produces small tissue movements in the plane of the push-pulse.These tissue movements,in turn,produce SWs that propagate and produce other minimal tissue movements in the horizontal plane of the push-pulse. These movements in the horizontal plane further propagate through the tissue in a sideway direction,away from the push-pulse. These SW movements (strictly speaking,tissue movements produced by SWs) are tracked by the regular-interval tracking US pulses,which are used to measure the arrival time of SWs. The simple formula is arrival time of SW/distance from the push-pulse,used to determine the speed of SWs. Such a measurement is possible because the speed of SWs is very slow (< 1-10 m/s)compared to the speed of US pulses (around 1540 m/s) (Table 1)[28,29]. There are two ways to quantify the relative tissue stiffness: SW speed expressed as m/s;and,kilopascal (kPa). Mostly,the SW speed is converted automatically to kPa,using the equation 1 kPa = 3 × P × (SW speed m/s)2,with the assumption that the examined tissue is always homogeneous,and where P is the tissue density and defined as 1.00 kg/m3.

    Table1 Acoustic characteristics of human tissues[28,29]

    The 2D-SWE method uses emission of multiple acoustic push-pulse to generate SWs from multiple lines in the tissue (Figure 1). As the SWs propagate,they cause tissue displacement,and this very minimal displacement is tracked by conventional pulse-echo US beams[6,30,31]. The US machine measures the propagation speed of SWs as they travel in tissue,and the measured SW speed at each point is displayed in color[6,32]. The reconstructed 2D color mapping of SWs is called “2D-SWE” and facilitates recognition of global distribution of tissue elasticity of the area.

    FACTORS THAT DEGRADE QUALITY OF TWODIMENSIONAL SHEAR WAVE ELASTOGRAPHY IMAGES

    For all,these US artifact-related problems are more clearly shown upon use of propagation mode,which can determine the causal artifact.

    Reverberation artifact

    Acoustic reverberation of the push-pulse at the liver capsule results in a false increase in SW speed. Reverberation artifacts arise when the US signals reflects repeatedly between two interfaces of largely different acoustic impedance,resulting in delayed echoes returning to the transducer at regular intervals. This can lead to reverberation artifacts on grayscale US image[33]. The same phenomenon of the push pulse is considered to occur between the liver capsule and the transducer. This phenomenon usually appears in the upper part of the cursor,presenting as a band of red (i.e. very increased SW speeds) (Figure 2). Thus,it is highly recommended to avoid this area when measuring SW values with SWE;if not,the results will be overestimated.

    Attenuation artifact

    Attenuation frequently occurs in a clinical setting and most commonly in obese or overweight patients. In these patients,a strongly attenuated push-pulse passing through the tissue results in poor creation of SWs at depth. This phenomenon usually appears at the distal part of the cursor,presenting as a colorless area (i.e. poor acquisition of SW data) (Figure 3). Thus,it is highly recommended to avoid this area when measuring SW values in obese or overweight patients.

    Probe compression artifact

    Excessive probe compression during image acquisition causes artificial compression of tissues just under the probe. This phenomenon is seen clearly when the liver is evaluated through subcostal scanning (Figure 4). SW measurement in the compressed area leads to a false elevation in SW values. A probe compression artifact appears near the probe,presenting as a reddish band. Thus,it is recommended not to compress the liver too much. The simplest way to avoid this artifact is to perform measurements through intercostal spaces only,where the liver tissue is less likely to be compressed by the probe.

    Motion artifact

    This phenomenon is seen clearly when the patient is unable to hold his/her breath or to change position quickly. Besides these extreme conditions,a motion artifact is usually seen when observing the left lobe of the liver,due to cardiac motion. Motion artifacts give rise to the characteristic image of a colorless area or a reddish area in the cursor,suggesting that SW values cannot be measured accurately in this area (Figure 5;see also,the section on tracking US beam-related problems below).

    Figure1 Schematic drawing of two-dimensional shear wave elastography in a healthy individual.

    MECHANISMS UNDERLYING ULTRASOUND ARTIFACTRELATED PROBLEMS IN TWO-DIMENSIONAL SHEAR WAVE ELASTOGRAPHY DETERMINED BY A COMPUTER SIMULATION MODEL

    US scanners reconstruct US images on the assumption that sound passes through all parts of the human body in a straight line and at a constant velocity,and apply this assumption to all scanning planes. Actually,however,these sound velocity in the human body varies with the composition of the tissue scanned. When a plane containing tissues with different velocities is scanned,sound refraction occurs at the interface of the tissues,according to Snell's law (see later in “Discussion”). Computer simulation analysis helps understand the global images of refraction in the plane,measuring the degree of refraction at each point of the interface,not only in grayscale US image but also in 2D-SWE image.

    2D-SWE in diffuse liver disease

    2D-SWE has been reported as useful for assessing liver fibrosis with high accuracy[8,9,34-36]. It is generally reported that SW speed increases as fibrosis advances[15,37-39];thus,heterogeneous distribution of different degrees of fibrosis will create a heterogeneous color mapping in 2D-SWE. Measurement of liver stiffness is based on the following assumptions: (1) All push-pulses emitted from the transducer advance directly without deviation,at an even distance from each other;and (2) All SWs created from these push-pulses propagate horizontally at the same distance from the transducer. However,in reality,like conventional US beams,push-pulses refract at the liver surface (according to Snell's law) because there is a large difference between the liver and the surrounding tissue and the hepatic surface is irregular.

    Figure6 shows the mode of propagation of push-pulses in the liver. As a result,SWs propagate in unassumed directions. This computer simulation model clearly teaches us an important lesson about measurement of SW speeds in liver cirrhosis.Furthermore,this speculation is in accordance with the findings from some studies with 2D-SWE measurements of healthy volunteers in whom the liver surface was smooth[40,41]. In those studies,measurements by novice operators were not significantly different from those of experienced operators[31,40,42]. Our computer simulation model shows that studies of normal liver do not contribute to resolving the problem of significant variability of SW values in liver cirrhosis. Although the measuring error is minimal,it is recommended to measure SW values through the least deformed hepatic surface possible.

    2D-SWE in liver tumors

    2D-SWE findings have been reported as useful for predicting tissue characterization of liver tumor,with markedly contradictory results[43-47]. The most widely accepted method for diagnosing liver tumors by 2D-SWE is measurement of maximal or median SW values within the lesion[46-49]. This method has long been performed but,as is shown in Figure 7,this idea is compromised by a combination of US refraction and reflection. The reported SW measurement of tumor stiffness is based on the following assumptions: (1) All SWs created from the push-pulses advance directly,without deviation,from the surrounding tissue toward the tumor and within the tumor;and(2) No reflection occurs at the interface between the surrounding tissue and liver tumor.

    Figure2 Reverberation artifact.

    However,in reality,like conventional US beams,SWs refract at the interface (again,according to Snell's law) because there is a large difference in propagation speed between the liver tumor and the surrounding tissue (Figure 7). Furthermore,the degree of refraction is thought to be more accentuated in 2D-SWE (1-10 m/s) than in conventional US (around 1540 m/s) (Figure 7B and C). The degree of reflection at the surrounding tissue-liver tumor interface is more accentuated in 2D-SWE,for the same reason. It is the deviation from the two above-mentioned assumptions that leads to the conclusion that accurate stiffness measurement of liver tumors is not possible by 2D-SWE,in most cases. This idea is in accordance with the American Institute of Ultrasound in Medicine recommendation that evaluation of liver tumor stiffness by 2D-SWE must be done with caution. Although,the SW value measurement inside the tumor is not highly accurate,this measurement is still valuable for judgement of relative stiffness of tumor and surrounding liver. However,conversely,this phenomenon helps in the detection of a liver tumor that is otherwise poorly visualized on B-mode (Figure 8) and in the differentiation of a pseudo-liver tumor(i.e.,focal fatty infiltration[43]or chronic hepatic porphyria[50]) from a true liver tumor.

    Tracking US beam-related problems

    Finally,we would like to briefly mention refraction of tracking US beams. We mentioned earlier the influence of US artifacts on 2D-SWE,on the assumption that the tracking US pulses occur at regular intervals and advance straight from the transducer through the tissue. However,they deviate in some instances. Figure 9 shows such a case. US beams deviate twice when passing through a focal lesion,with one at the upper surrounding tissue-focal lesion interface and the other at the lower focal lesionsurrounding tissue interface. As a result,US beams propagating in tissue posterior to the focal lesion deviate;however,the US machine calculates SW values on the assumption that there is no deviation of US beams. Thus,this phenomenon suggests that SW measurements must be done in a plane that does not contain focal lesions.

    DISCUSSION

    In many reported studies,2D-SWE has been shown to have a sufficient level of sensitivity in evaluating the degree of hepatic fibrosis[7-9,36]but with significant heterogeneity of results. This heterogeneity has been considered due to variations among patient populations,study design,and devices used[49]. Some of these studies have pointed to technical flaws in earlier studies but recent studies also show some instable results[45,46,51]. These studies have brought into question whether they included the influence of US artifacts.

    Figure3 Attenuation artifact.

    This review has an important strength in that it is a theoretical analysis not found in similar trials reported in the literature. Computer simulation model yields a purely theoretical analysis. In the case of gray-scale US images,the sound refraction produces artifactual images,because the US scanner displays each point at the appropriate distance determined by the time taken for the echo to return to the transducer in the direction in which the transducer is pointing at the time,even when the US beam is refracted. In the case of 2D-SWE as well,the US scanner reconstructs 2D-SWE images on the assumption that SW passes horizontally in a straight line,without deviation,and the US scanner displays SW velocity mapping on the basis of data measured by tracking pulses (determined by the time/distance). Computer simulation model enables us to calculate accurately and automatically the degree of refraction at each point,and understand the global image of refraction in the plane.This method is especially useful for understanding the global mode of refraction at the curved interface (tumor-surrounding tissue interface or surrounding tissueirregular hepatic surface interface) (Figure 10)[52-54]. In short,it ensures that unfavorable factors,such as technical errors,differences in US machines used and different levels of 2D-SWE experience,or influence biased by additional clinical data,do not interfere in this analysis. It also maintains a high quality of interpretation. The SWE diagnosis is usually the so-called “final report” of the clinician and explains that it is necessarily correct. Our simulation model can be considered the basic thought of what the clinicians conclude. It is recommended,therefore,that as more practitioners are trained in 2D-SWE,our basic thought is used in clinical practice.

    CONCLUSION

    Two-dimensional shear wave elastography artifacts,although seen very frequently in a clinical setting,are poorly recognized. Our review emphasizes that interpretation of 2D-SWE images must incorporate knowledge of US artifacts.

    Figure4 Probe compression artifact.

    Figure5 Motion artifact.

    Figure6 Computer simulation model of push-pulses in a cirrhotic patient.

    Figure7 Two-dimensional shear wave elastography of liver tumor.

    Figure8 Representative case: intrahepatic cholangiocellular carcinoma.

    Figure9 Computer simulation model of deviated push-pulses.

    Figure10 Sound refraction at the interface.

    99国产精品一区二区蜜桃av| 欧美高清成人免费视频www| 少妇的丰满在线观看| 他把我摸到了高潮在线观看| 欧美又色又爽又黄视频| 两个人的视频大全免费| 亚洲中文字幕日韩| 亚洲熟女毛片儿| 欧美日韩综合久久久久久 | 欧美性猛交黑人性爽| 宅男免费午夜| 亚洲,欧美精品.| 欧美一区二区国产精品久久精品| 一二三四社区在线视频社区8| 国产成人福利小说| 深夜精品福利| 日韩欧美一区二区三区在线观看| 天天躁狠狠躁夜夜躁狠狠躁| 欧美日韩黄片免| 我的老师免费观看完整版| 亚洲熟女毛片儿| 九九在线视频观看精品| 激情在线观看视频在线高清| 亚洲av五月六月丁香网| 国产黄片美女视频| 久久香蕉精品热| 人妻夜夜爽99麻豆av| 免费在线观看影片大全网站| 人人妻人人澡欧美一区二区| 久久久久久国产a免费观看| 免费观看的影片在线观看| 99久久成人亚洲精品观看| 激情在线观看视频在线高清| 久久亚洲精品不卡| www.www免费av| 日韩欧美在线二视频| 两性午夜刺激爽爽歪歪视频在线观看| 国产真人三级小视频在线观看| 99国产综合亚洲精品| 欧美日韩福利视频一区二区| 国产探花在线观看一区二区| 日本免费一区二区三区高清不卡| 色综合站精品国产| 午夜激情欧美在线| 香蕉久久夜色| 日韩欧美精品v在线| 99国产综合亚洲精品| 亚洲欧美激情综合另类| 亚洲欧美日韩东京热| 亚洲乱码一区二区免费版| 亚洲无线在线观看| 欧美丝袜亚洲另类 | 香蕉丝袜av| 97超级碰碰碰精品色视频在线观看| 看黄色毛片网站| www.自偷自拍.com| 亚洲九九香蕉| 熟女少妇亚洲综合色aaa.| 亚洲国产精品合色在线| 亚洲美女视频黄频| 禁无遮挡网站| 天堂av国产一区二区熟女人妻| 每晚都被弄得嗷嗷叫到高潮| 久久香蕉国产精品| aaaaa片日本免费| 亚洲成av人片在线播放无| 日韩人妻高清精品专区| 极品教师在线免费播放| 人妻久久中文字幕网| 亚洲欧美日韩东京热| 美女午夜性视频免费| 18禁美女被吸乳视频| 老司机午夜福利在线观看视频| 午夜福利18| 长腿黑丝高跟| 午夜激情欧美在线| 欧美精品啪啪一区二区三区| 精品国产美女av久久久久小说| 亚洲精品久久国产高清桃花| 脱女人内裤的视频| 黄色 视频免费看| 九色成人免费人妻av| 成年人黄色毛片网站| 精品国产乱子伦一区二区三区| 久久精品91蜜桃| 岛国视频午夜一区免费看| 真人一进一出gif抽搐免费| 欧美最黄视频在线播放免费| 人人妻人人澡欧美一区二区| 亚洲欧美日韩卡通动漫| 黑人欧美特级aaaaaa片| 国内少妇人妻偷人精品xxx网站 | 国产精品98久久久久久宅男小说| 国产成年人精品一区二区| 亚洲色图 男人天堂 中文字幕| 看免费av毛片| 日韩人妻高清精品专区| 91在线精品国自产拍蜜月 | 欧美日韩瑟瑟在线播放| 国产欧美日韩一区二区三| 国产亚洲av嫩草精品影院| 精品国产乱子伦一区二区三区| ponron亚洲| 午夜福利免费观看在线| 精品日产1卡2卡| 国产黄片美女视频| 深夜精品福利| 国产亚洲欧美98| 欧美绝顶高潮抽搐喷水| 午夜免费观看网址| 一本一本综合久久| netflix在线观看网站| 国产野战对白在线观看| 亚洲aⅴ乱码一区二区在线播放| 欧美日韩亚洲国产一区二区在线观看| 少妇的逼水好多| 真人做人爱边吃奶动态| 夜夜夜夜夜久久久久| 成年女人毛片免费观看观看9| 91老司机精品| 黄色日韩在线| 神马国产精品三级电影在线观看| 在线视频色国产色| 波多野结衣高清作品| 国产高清有码在线观看视频| 夜夜爽天天搞| 美女cb高潮喷水在线观看 | 亚洲va日本ⅴa欧美va伊人久久| 亚洲 欧美 日韩 在线 免费| 久久久久亚洲av毛片大全| 男人舔女人下体高潮全视频| 亚洲精品粉嫩美女一区| 成人特级av手机在线观看| 日日干狠狠操夜夜爽| 在线观看日韩欧美| 午夜成年电影在线免费观看| 美女被艹到高潮喷水动态| 精品不卡国产一区二区三区| 99re在线观看精品视频| 一级作爱视频免费观看| 亚洲国产日韩欧美精品在线观看 | 国产黄色小视频在线观看| 精品99又大又爽又粗少妇毛片 | 午夜激情福利司机影院| 一个人看的www免费观看视频| 欧美午夜高清在线| 无人区码免费观看不卡| 国产精品一区二区精品视频观看| 床上黄色一级片| 美女高潮的动态| 国产精品1区2区在线观看.| 亚洲av美国av| 少妇的丰满在线观看| 精品不卡国产一区二区三区| 俄罗斯特黄特色一大片| 真实男女啪啪啪动态图| 精品久久久久久,| a级毛片在线看网站| 亚洲av成人精品一区久久| 男人舔女人的私密视频| a级毛片在线看网站| 首页视频小说图片口味搜索| 亚洲午夜精品一区,二区,三区| 久久中文看片网| 三级毛片av免费| 757午夜福利合集在线观看| 97超视频在线观看视频| www国产在线视频色| 精品不卡国产一区二区三区| 免费看美女性在线毛片视频| 天堂网av新在线| 午夜福利在线在线| 啪啪无遮挡十八禁网站| 色综合婷婷激情| 久久亚洲精品不卡| 国内精品久久久久久久电影| 日韩中文字幕欧美一区二区| 天天添夜夜摸| 亚洲人成网站高清观看| 在线观看一区二区三区| 精品一区二区三区四区五区乱码| 国产乱人伦免费视频| 九色成人免费人妻av| 99久国产av精品| 国产精品1区2区在线观看.| 床上黄色一级片| 色综合欧美亚洲国产小说| 亚洲电影在线观看av| 综合色av麻豆| 国产精品一区二区免费欧美| 亚洲成人精品中文字幕电影| 久久久色成人| 日韩精品中文字幕看吧| АⅤ资源中文在线天堂| 久久草成人影院| 亚洲精品在线美女| 国产精品av久久久久免费| 国产精品久久电影中文字幕| 亚洲真实伦在线观看| 俺也久久电影网| 亚洲成av人片在线播放无| 日韩欧美一区二区三区在线观看| 久久久久性生活片| 99久久久亚洲精品蜜臀av| а√天堂www在线а√下载| 两个人看的免费小视频| 男女下面进入的视频免费午夜| 1024手机看黄色片| 三级男女做爰猛烈吃奶摸视频| 视频区欧美日本亚洲| 视频区欧美日本亚洲| 婷婷丁香在线五月| 午夜福利高清视频| 丝袜人妻中文字幕| 亚洲欧美日韩东京热| 国产av不卡久久| 性欧美人与动物交配| www日本在线高清视频| 12—13女人毛片做爰片一| 欧美黄色淫秽网站| 免费在线观看日本一区| 午夜视频精品福利| 嫩草影院精品99| 成熟少妇高潮喷水视频| 日本成人三级电影网站| 在线视频色国产色| 久久天躁狠狠躁夜夜2o2o| 欧美大码av| 国产乱人视频| 偷拍熟女少妇极品色| 亚洲av电影不卡..在线观看| 免费在线观看日本一区| 免费av不卡在线播放| 中文字幕人成人乱码亚洲影| 综合色av麻豆| aaaaa片日本免费| 午夜日韩欧美国产| 亚洲av成人不卡在线观看播放网| av欧美777| 亚洲18禁久久av| 国产成人影院久久av| 成人18禁在线播放| 免费在线观看日本一区| 亚洲最大成人中文| 日韩中文字幕欧美一区二区| 午夜福利在线在线| 黄色视频,在线免费观看| 国产视频内射| 又黄又粗又硬又大视频| 国产精品久久久av美女十八| 精品熟女少妇八av免费久了| 日韩中文字幕欧美一区二区| 国产精品久久视频播放| 欧美绝顶高潮抽搐喷水| 亚洲性夜色夜夜综合| 淫妇啪啪啪对白视频| 亚洲国产看品久久| 最新在线观看一区二区三区| 国产精品久久久久久亚洲av鲁大| 国产淫片久久久久久久久 | 欧美成狂野欧美在线观看| 亚洲欧美日韩东京热| 亚洲av日韩精品久久久久久密| 成人18禁在线播放| 一区福利在线观看| 免费看日本二区| 一进一出抽搐gif免费好疼| 麻豆av在线久日| 欧美乱码精品一区二区三区| 国产一区二区激情短视频| 国产精品永久免费网站| 99国产极品粉嫩在线观看| 欧美成人性av电影在线观看| 男女那种视频在线观看| 色尼玛亚洲综合影院| 精品一区二区三区四区五区乱码| 日韩欧美在线乱码| 色吧在线观看| 中文字幕精品亚洲无线码一区| 免费无遮挡裸体视频| 国产一区二区激情短视频| 国产97色在线日韩免费| 日本成人三级电影网站| 久久精品影院6| av片东京热男人的天堂| 国产伦一二天堂av在线观看| 亚洲,欧美精品.| 色视频www国产| 亚洲成人免费电影在线观看| 午夜福利免费观看在线| 国产综合懂色| 亚洲乱码一区二区免费版| 这个男人来自地球电影免费观看| 中文字幕久久专区| 婷婷六月久久综合丁香| xxxwww97欧美| 国产亚洲精品久久久久久毛片| 偷拍熟女少妇极品色| 亚洲av成人精品一区久久| 18禁黄网站禁片免费观看直播| 欧美乱妇无乱码| 老熟妇乱子伦视频在线观看| 成年女人永久免费观看视频| 校园春色视频在线观看| www.自偷自拍.com| 这个男人来自地球电影免费观看| 丁香六月欧美| 啦啦啦免费观看视频1| 久久性视频一级片| 免费在线观看视频国产中文字幕亚洲| 岛国视频午夜一区免费看| 中出人妻视频一区二区| 国产极品精品免费视频能看的| a级毛片a级免费在线| 伦理电影免费视频| 波多野结衣高清无吗| 亚洲精品久久国产高清桃花| 亚洲成av人片在线播放无| 色综合亚洲欧美另类图片| 老司机午夜十八禁免费视频| 国产精品98久久久久久宅男小说| 视频区欧美日本亚洲| 欧美国产日韩亚洲一区| 婷婷精品国产亚洲av在线| 亚洲国产欧美一区二区综合| 亚洲 欧美一区二区三区| 国产高清激情床上av| 在线免费观看不下载黄p国产 | 一二三四在线观看免费中文在| 成年女人永久免费观看视频| 国产久久久一区二区三区| 又粗又爽又猛毛片免费看| 亚洲国产欧美一区二区综合| 国产成人精品久久二区二区免费| 久久精品aⅴ一区二区三区四区| bbb黄色大片| 亚洲专区字幕在线| 久久精品国产亚洲av香蕉五月| 美女cb高潮喷水在线观看 | 久久久久久大精品| 中出人妻视频一区二区| 全区人妻精品视频| 久久久久久国产a免费观看| 美女高潮的动态| 久久国产精品影院| 国产亚洲欧美在线一区二区| 欧美一级毛片孕妇| 亚洲 欧美 日韩 在线 免费| 久久久精品欧美日韩精品| 法律面前人人平等表现在哪些方面| 久久精品91蜜桃| 国产精品av久久久久免费| 在线播放国产精品三级| 亚洲av成人不卡在线观看播放网| 国产精品久久视频播放| 国产成人精品久久二区二区免费| 色综合亚洲欧美另类图片| 精品午夜福利视频在线观看一区| 久久人人精品亚洲av| 在线免费观看不下载黄p国产 | 亚洲av电影不卡..在线观看| 一区福利在线观看| 天堂影院成人在线观看| 日本精品一区二区三区蜜桃| 国产激情欧美一区二区| 淫秽高清视频在线观看| 91九色精品人成在线观看| 天天添夜夜摸| 老司机深夜福利视频在线观看| 俄罗斯特黄特色一大片| 国产精品乱码一区二三区的特点| 91久久精品国产一区二区成人 | 精品一区二区三区视频在线观看免费| 欧美乱妇无乱码| 国产极品精品免费视频能看的| 国产精华一区二区三区| 两个人的视频大全免费| 午夜激情欧美在线| 无遮挡黄片免费观看| 色视频www国产| 村上凉子中文字幕在线| 91麻豆av在线| 99久久精品热视频| 黑人欧美特级aaaaaa片| 麻豆国产97在线/欧美| 欧美最黄视频在线播放免费| 久久久久久大精品| 特级一级黄色大片| 夜夜爽天天搞| 日本黄色视频三级网站网址| 成人av在线播放网站| 男人舔女人的私密视频| 国产亚洲欧美98| 久久国产乱子伦精品免费另类| 美女高潮的动态| 97人妻精品一区二区三区麻豆| 久久精品国产99精品国产亚洲性色| 91av网一区二区| 欧美一级毛片孕妇| 色综合亚洲欧美另类图片| 国产综合懂色| 国产成人福利小说| 久久久久免费精品人妻一区二区| 国产黄色小视频在线观看| 在线十欧美十亚洲十日本专区| 91久久精品国产一区二区成人 | 一进一出抽搐gif免费好疼| 亚洲成人久久爱视频| 神马国产精品三级电影在线观看| 狂野欧美激情性xxxx| 国产真人三级小视频在线观看| 国产精品久久电影中文字幕| 97超视频在线观看视频| 亚洲av中文字字幕乱码综合| 叶爱在线成人免费视频播放| 欧美不卡视频在线免费观看| 黄片大片在线免费观看| 国产爱豆传媒在线观看| 首页视频小说图片口味搜索| 久久这里只有精品中国| 黑人欧美特级aaaaaa片| 久久精品综合一区二区三区| 国产免费男女视频| 麻豆av在线久日| 精品一区二区三区视频在线观看免费| 女警被强在线播放| 成人三级黄色视频| 丰满人妻熟妇乱又伦精品不卡| 在线免费观看的www视频| 国产探花在线观看一区二区| 欧美性猛交╳xxx乱大交人| 特级一级黄色大片| 国模一区二区三区四区视频 | 国产亚洲欧美98| 美女午夜性视频免费| 国内久久婷婷六月综合欲色啪| 欧美午夜高清在线| 丁香六月欧美| 亚洲av第一区精品v没综合| 国产毛片a区久久久久| 嫩草影院入口| 人妻夜夜爽99麻豆av| 在线免费观看不下载黄p国产 | 久久午夜综合久久蜜桃| 国产精品九九99| 好看av亚洲va欧美ⅴa在| 两性夫妻黄色片| 日本黄大片高清| 在线观看免费午夜福利视频| 国产三级在线视频| 亚洲国产精品久久男人天堂| 国产欧美日韩精品亚洲av| 国产亚洲精品综合一区在线观看| а√天堂www在线а√下载| 变态另类成人亚洲欧美熟女| 一级作爱视频免费观看| 日本与韩国留学比较| 精品久久久久久成人av| 国产精品一区二区三区四区久久| 黑人操中国人逼视频| а√天堂www在线а√下载| 在线观看66精品国产| 麻豆国产av国片精品| 亚洲av成人不卡在线观看播放网| 午夜福利成人在线免费观看| 88av欧美| 好看av亚洲va欧美ⅴa在| 久久亚洲真实| 我的老师免费观看完整版| xxxwww97欧美| 久久久成人免费电影| 脱女人内裤的视频| 性色av乱码一区二区三区2| 亚洲精品一卡2卡三卡4卡5卡| 真人一进一出gif抽搐免费| 午夜免费成人在线视频| 法律面前人人平等表现在哪些方面| 亚洲18禁久久av| 麻豆一二三区av精品| 日本a在线网址| 三级男女做爰猛烈吃奶摸视频| 欧美乱妇无乱码| 国产极品精品免费视频能看的| 一本久久中文字幕| 男人舔奶头视频| 美女扒开内裤让男人捅视频| www.999成人在线观看| 国产亚洲精品一区二区www| 亚洲国产欧美一区二区综合| 午夜福利高清视频| 欧美另类亚洲清纯唯美| 国产欧美日韩精品亚洲av| 一本综合久久免费| 一本久久中文字幕| 亚洲精品一区av在线观看| 亚洲精品乱码久久久v下载方式 | 非洲黑人性xxxx精品又粗又长| 国产又色又爽无遮挡免费看| 99热精品在线国产| 成人午夜高清在线视频| 我的老师免费观看完整版| 国产成人欧美在线观看| 亚洲av美国av| 久久精品亚洲精品国产色婷小说| 嫁个100分男人电影在线观看| 伊人久久大香线蕉亚洲五| 欧美另类亚洲清纯唯美| 麻豆国产97在线/欧美| 岛国在线观看网站| 麻豆成人午夜福利视频| 日韩 欧美 亚洲 中文字幕| 亚洲欧美日韩高清在线视频| 性色av乱码一区二区三区2| 精品国产亚洲在线| 国产单亲对白刺激| 深夜精品福利| 免费在线观看视频国产中文字幕亚洲| 国产亚洲精品久久久com| 亚洲精品一区av在线观看| 级片在线观看| 舔av片在线| 日本成人三级电影网站| 欧美色欧美亚洲另类二区| 一进一出好大好爽视频| 韩国av一区二区三区四区| 欧美成人一区二区免费高清观看 | 老司机福利观看| 长腿黑丝高跟| 国产精品一区二区三区四区免费观看 | 十八禁网站免费在线| 国产精品99久久久久久久久| 久久午夜亚洲精品久久| 97人妻精品一区二区三区麻豆| 亚洲在线观看片| 亚洲精品中文字幕一二三四区| 色哟哟哟哟哟哟| 色综合欧美亚洲国产小说| 亚洲熟妇熟女久久| 曰老女人黄片| 给我免费播放毛片高清在线观看| 日韩欧美 国产精品| 无人区码免费观看不卡| 亚洲成人中文字幕在线播放| 国产亚洲欧美在线一区二区| 99热精品在线国产| 听说在线观看完整版免费高清| 久久午夜综合久久蜜桃| 久久精品综合一区二区三区| 欧美成人免费av一区二区三区| 亚洲自拍偷在线| 少妇的丰满在线观看| 一级a爱片免费观看的视频| 亚洲男人的天堂狠狠| 午夜免费观看网址| 精品久久久久久久久久免费视频| 国产成年人精品一区二区| 国产精品电影一区二区三区| 综合色av麻豆| 欧美日韩瑟瑟在线播放| 午夜免费激情av| 国产视频一区二区在线看| 精品一区二区三区视频在线 | 欧美日韩黄片免| 欧美日韩福利视频一区二区| 999久久久国产精品视频| 亚洲成a人片在线一区二区| 国产野战对白在线观看| 色综合欧美亚洲国产小说| 他把我摸到了高潮在线观看| 精品一区二区三区av网在线观看| 国产亚洲欧美在线一区二区| 亚洲国产日韩欧美精品在线观看 | 午夜福利视频1000在线观看| 999久久久国产精品视频| 久久久国产成人精品二区| 一区福利在线观看| 亚洲人成网站高清观看| 久久久国产精品麻豆| 成人永久免费在线观看视频| 亚洲黑人精品在线| 精品一区二区三区视频在线观看免费| 熟女人妻精品中文字幕| 国产野战对白在线观看| 美女扒开内裤让男人捅视频| 在线永久观看黄色视频| 欧美中文综合在线视频| 亚洲成人中文字幕在线播放| www.熟女人妻精品国产| 男女床上黄色一级片免费看| 久久精品国产综合久久久| 免费人成视频x8x8入口观看| 一区二区三区高清视频在线| 黄色丝袜av网址大全| 观看美女的网站| 久久久久久九九精品二区国产| 亚洲第一电影网av| 偷拍熟女少妇极品色| 日韩欧美 国产精品| 伦理电影免费视频| 亚洲精品中文字幕一二三四区| 久久精品人妻少妇| 亚洲人成网站高清观看| 啪啪无遮挡十八禁网站| 男女午夜视频在线观看| 国产激情欧美一区二区| 十八禁人妻一区二区| 美女午夜性视频免费| 9191精品国产免费久久| 毛片女人毛片| 亚洲美女视频黄频| 搞女人的毛片| 亚洲人与动物交配视频| 久久中文看片网| 亚洲av五月六月丁香网| 国产69精品久久久久777片 |