    An Implanted Closed-loop Chip System for Heart Rate Control:System Design and Effects in Conscious Rats☆

    2010-11-02 07:27:10YuxunZhouYunYunJunGoLingYngFengZhngGuoqingZhuXingyGo
    THE JOURNAL OF BIOMEDICAL RESEARCH 2010年2期

    Yuxun Zhou, Yun Yun, Jun Go, Ling Yng, Feng Zhng, Guoqing Zhu, Xingy Go*

    aDepartment of Biomedical Engineering, Nanjing Medical University, Nanjing, 210029, China

    bDepartment of Physiology, Nanjing Medical University, Nanjing, 210029, China

    INTRODUCTION

    Medication is most commonly used to treat such cardiovascular conditions as hypertension, heart failure, and arrhythmias. However, these drugs have many side effects. For example, the conventional medication for hypertension may lead to orthostatic hypotension, arrhythmias and water and electrolyte disorders[1]. Using nitrates to treat heart failure may result in headache, hypotension and dizziness[2].Resistance may develop to many of these drugs.Furthermore, almost all of them need to be taken for life and this is hard for most patients to accept. A nonpharmacotherapy method to solve these problems is highly desirable.

    Implantable devices provide a feasible alternative in such cases[3-6]. Instead of taking medicine every day, the patients may prefer receiving an implanted device to relieve the symptoms for a relatively long time. Actually the developments in therapeutic and diagnostic devices have increasingly improved the patients’ quality of life. The rapid development of microelectronics technology improves the possibility of the clinical application of implantable devices.The implantable device has become a research focus in the biomedical field, and both medical researchers and physicians are realizing that such devices may be a new therapy for some disorders that are resistant to drug therapy.

    Tachycardia is a common clinical arrhythmia, and vagus nerve stimulation (VNS)may be a potential way to cure it. VNS has been proposed as a possible treatment in several conditions, e.g. myocardial infarction (MI)[7], atrial fibrillation (AF)[8]and angina pectoris[9]. It has also proved to be an effective way to influence heart rate in both animals and humans[10-11]. An implanted vagus nerve stimulator could work in synergy with β-blockers and ACE inhibitors, by enhancing their life-saving action[12]. However, the human body is a complex closed-loop regulation system.In patients with dysfunctional control mechanisms,the general open-loop device is not safe or effective when interacting with the circulatory system[4,5,13].Therefore, an Implanted Closed-loop Chip System was developed to control the circulatory function and cure rapid arrhythmias. In our past studies the implanted chip system has been successful in controlling blood pressure in both SD rats and spontaneously hypertensive rats (SHR)[4,5,14]. In order to regulate the HR smoothly and safely, a similar implantable closed-loop chip system was designed.The chip system function included collecting heart rate signals, performing data analysis, providing vagus nerve stimulation and maintaining external wireless communication. The purpose of the present study was to employ the chip in vivo, and to determine its efficiency in regulating heart rate in normal conscious rats.

    MATERIALS AND METHODS

    System Description

    The whole system consisted of the slave machine(ECG sensing electrodes, stimulation electrodes,amplifiers, a processor, a wireless transceiver and a silver-zinc battery), the master machine (processor,a wireless-transceiver and a USB bridge circuit)and PC-based user software (for displaying ECG signals and controlling the slave machine). The slave machine detected the ECG signals from the two precordial electrodes. The signals were amplified with a custom amplifier and fed to an on-chip 10-bit A/D convertor sampling at 500 Hz. The digital ECG data were sent to the master machine by the Nordic wireless transceiver NRF24L01. On the master side, another wireless transceiver collected the signals and sent back commands by way of half duplex communication. A MSP430 processor was used for buffering the data,formatting the chip, and communicating with a PC by a USB bridge circuit (Fig. 1). The slave machine was implanted subcutaneously in the abdomen. The stimulation electrodes were placed around the right cervical vagus nerve, and the precordial recording electrodes were located subcutaneously. The ECG was detected and analyzed by the chip. Depending on the instantaneous ECG R-R intervals, the system generated corresponding frequency electric pulses to stimulate the vagus nerve. At the same time through wireless transmission, host computer software received and displayed the real-time ECG waveforms and cardiac cycle. In addition, 24-h data was also stored in a flash memory of the chip.

    The Slave Machine

    The Processor

    A minitype processor was used as the MPU of the slave machine, which allowed the chip to be small enough to be implanted in a rat. The processor had two timers: Timer A and a Watchdog Timer. Timer A was assigned to control the frequency and stimulation pulse width, while the Watchdog Timer contributed to the sampling rate of the A/D convertor for ECG signal collecting[15,16]. The real time results of the A/D convertor were sent wirelessly through the Nordic NRF24L01 to the master machine and the average cardiac cycles were saved to the flash memory at the same time. Thus the intact heart rate changing trend could be conserved during the experiment to avoid some accidental wireless communication corruptions.Almost 30 h of data were stored in the flash memory of the slave machine and could be collected afterwards through a user command by the master machine.The Amplifier

    In order to achieve a minimum component count and smaller size, we used one rail-to-rail I/O dual Operational Amplifier (OPA2369, Texas Instruments,USA)to build a differential amplifier (Fig. 2). The output voltage of the circuit could be calculated by so the gain here was 54 dB. A voltage division circuit composed of R11, R12and C3was designed to raise the baseline of the amplifier output in order to fit the onchip A/D convertor. The pass-band of the circuit was 1.6 Hz-40 Hz.

    R-R Interval Recognition

    As a closed-loop system, stimulation could be dynamically adjusted according to the real-time heart rate (HR). Continuously acquired ECG data were used to recognize the R-R intervals. The absolute value of 9-point difference data was compared with the threshold. When the result was positive, an R wave was recognized. Still while an R wave was recognized,a refractory timer was set. Thus, during the refractory period no other difference data would be recognized as an R wave until the timer ended in order to avoid repeated error recognition of the same R wave. The refractory timer could be set according to the different experimental animals. Obviously the real-time HR could be calculated by the measurement of the cardiac cycle.

    Fig. 1 System diagram. a:Recording ECG, HR; b:ECG electrodes; c.stimulation.

    Fig. 2 ECG amplifier schematics

    Stimulation Method

    The stimulation part of the system was controlled by the PWM generator of the chip. As we intended to produce a 6V stimulation with a 3V power supply, the circuit was designed as follows. A differential voltage of 6V was achieved due to the 2 PWM outputs (Fig. 3). As illustrated in Fig. 4, when a pulse occurred one of the PWM pulses was +3V and the other one was -3V (after the capacitor), hence the differential voltage between the two stimulation electrodes reached 6V. The two capacitors ensured the voltage became zero after the pulse so that there was no any direct current between the two electrodes, avoiding tissue electrolysis.

    The basic regulation principle of the chip system is simply expressed as the relationship between HR and frequency of stimulus. When the HR was between 300 and 580 bpm, the stimulus frequency increased following the elevation of the HR. The pattern of stimulus changing is represented in Fig. 4. The increased stimulation of the vagus nerve consequently caused a decrease in the HR. When the HR exceeded 580 bpm, the frequency of stimulus reached its maximal value of 25 Hz. However when the HR was less than 300 bpm, the stimulation stops automatically(Fig. 5). In addition, the upper limit and lower limit of the relationship between stimulation and HR could be set by the master machine through the wireless communication.

    The Master Machine

    The master machine served as a data bridge between the slave machine and PC. It received data from the slave machine and then sent them to the PC user software. At the same time the user’s command would be sent back to the slave machine. The user command included:

    ① Frequency of stimulation control;

    ② Stimulation starting point control (upper and lower limits of the HR stimulation relationship);

    ③ R wave recognition threshold control;

    Fig. 3 PWM stimulation schematics.

    Fig. 4 The pattern of stimulus changing.

    Fig. 5 Relationship between HR and the frequency of stimulation in the chip systems.

    ④ PWM stimulation duty cycle control;

    ⑤ Temporary data sending (the temporary data stored in the flash memory of the slave machine);

    ⑥ Temporary data clearance.

    The RF link data were collected via a Nordic nRF24L01 transceiver on the master side. The transceiver was operated by a MSP430F169 microcontroller and the communication between the master machine and PC achieved by a USB bridge IC(FTDI FT245BM).

    Wireless Communication

    The Nordic nRF24L01 was chosen to deal with the wireless communication. It was a single chip 2.4 GHz transceiver with an embedded baseband protocol engine (Enhanced ShockBurstTM), designed for ultra low power wireless applications.

    The main features of Enhanced ShockBurstTMare as follow:

    ① 1 to 32 bytes dynamic payload length;

    ② Automatic packet handling;

    ③ Auto packet transaction handling;

    ④ Auto acknowledgement;

    ⑤ Auto retransmit;

    ⑥ 6 data pipe MultiCeiverTMfor 1:6 star networks

    Importantly, the dynamic payload mode feature could achieve the half duplex communication between the slave and the master side. Therefore,without changing the mode from sending to receiving, the power consumption was ultra low during two-way communication. In addition, with the auto packet transaction handling feature the communication was quite safe and robust. One master machine could control 6 slave machines, thus the data from 6 different test animals could be collected simultaneously[17].

    User Software

    The whole system was configured with the software MD2000WL running on Windows-based PC’s. The software of the MD2000WL data acquisition system(Nanjing Medical University)was developed by MFC,including data plotting, data filtering and data analysis.MD2000WL was based on the ordinary MD2000U system with the addition of the slave machine setting and control interface. The control function included stimulation starting point control (upper and lower limits of the heart rate stimulation relationship), flash data control, R wave recognition threshold control,and stimulation duty cycle control.

    Animals and Surgical procedure

    Six male SD rats with normal HR weighing 400-500 g were used to test the system. All animals were obtained from Laboratory Animal Centers of Jiangsu Province and were handled according to the guidelines approved by the Experimental Animal Care and Use Committee of the Nanjing Medical University. Each rat was anesthetized with an intraperitoneal injection of chloral hydrate (0.4 g/kg). The right cervical vagus nerve was isolated and a pair of silver electrodes was placed around the nerve in preparation for the stimulations. Two silver detection electrodes were placed under the skin of the chest to acquire ECG signals with one at the cardiac apex and the other at the cardiac base. The main part of the chip was subcutaneously implanted into the abdomen with wires tunneled under the skin to connect with the stimulation electrodes and the detection electrodes.The rats were caged in a controlled temperature and humidity environment with a 12 h light/dark cycle.Standard laboratory chow and drinking water were available ad libitum.

    Protocols

    1-3 h after the surgical process, rats recovered from anesthesia. We spent an additional 2 h allowing the animals to resume normal behavior, after which each rat was subjected to a 30-min regulation and 30-min recovery period. ECG monitoring lasted for 24 h after the surgery in order to test if the chip system and its wireless communication were dependable.

    Statistical Analysis

    Differences among the three observations in the same animal were assessed by Student’s paired t test using SPSS 16.0. The criterion for statistical significance was set at P < 0.05. All data are presented as mean ± SD.

    RESULTS

    ECG signals were successfully collected during the experiment. R-R interval recognition accuracy was more than 95%. The restlessness of the rats gave rise to most errors in R-R interval recognition. An example of the ECG signals acquired is shown in Fig. 6.

    The HR was controlled successfully by the chip system. The change in HR during regulation and recovery is shown in Fig. 7. During the regulation period, the HR was stably controlled at a lower level after an initial stage of fluctuation. After the regulation the stimulation was stopped and the HR returned to baseline in 1 min.

    HR Changes

    The HR decreased significantly during the period of regulation compared with the control period(-79.3±34.5, P < 0.01, n = 6)and with the recovery period (-79. 0±28.8, P < 0.01, n = 6). There was no significant difference in HR between the control and the recovery periods (Fig. 7B).

    Fig. 6 Representative ECG recording in conscious rats after the surgery

    DISCUSSION

    Fig. 7 Representative changes (A)and statistical analysis (B)of HR recording during the periods of regulation and the recovery. A:representative changs of HR; B: statistical analysis of HR, ** P < 0.01.

    The primary finding of the present study was that a closed-loop implanted chip system used to control HR was successfully established in conscious animals.There is ample evidence that direct electrical stimulation of the vagus nerve can change HR in both animals and humans[10-11]. However, the use of an implanted chip to regulate HR at predetermined levels has not yet been reported. This study presents an example of using a closed-loop implanted chip system to artificially regulate the physiological function of an animal. We have successfully produced a Remote-controlled animal model for the control of visceral functions. It is more important to control HR in conscious rats than in anesthetized animals.

    The chip system was based on the theory of establishing an artificial set point to take the place of the animal's own set point. ECG signals were recorded and converted to HR data through R-R interval recognition, and then the signals were sent back to the animals by means of stimulating the cervical vagus nerve. Whenever HR exceeded the set point, the stimulation to the vagus nerve would start. The higher the HR went, the stronger the stimulation generated.In this way, the closed-loop regulation maintained HR in a relatively narrow range, which was slightly higher than the artificial set point. It was found that there was a damped oscillation of HR after the regulation started, which might indicate that the chip regulation acted as a moderately damped negative feedback loop.When HR accelerated, the stimulation would become stronger as programmed, thereby reducing the HR.Until the regulation system of the body adapted to the chip system, HR changing became stable. Because of the complexity of the HR regulating mechanism,it is still not fully understood whether slowing of HR is mediated by activation of efferent vagus fibers or by sympathetic reflex inhibition through afferent fibers[10,18].The detailed physiological mechanism of such a damped oscillation is not well known.

    The goal of closed-loop control of HR has been achieved and the system worked steadily without any unexpected problems. Nevertheless, improvement is required. Firstly, in contrast to anaesthetized animals, after such a surgery the conscious animals were agitated because of the pain and fear and their restlessness sometimes influenced our data collection. Although the R-R interval recognition was basically accurate because of the algorithm used, the electrodes fixation method and the ECG collection pattern still need to be improved. Secondly the size of the slave machine should be reduced in the future with a specific integrated circuit design so as to not overburden the test animals.

    In summary, the chip system successfully regulated HR to an artificial set point in conscious rats, and this set point could be changed remotely. The present study laid a foundation for using the chip system to regulate physiological function and potentially to treat disease.

    [1]Kario K, Shimada K, Pickering TG. Pressor effects of beta-blockers on standing blood pressure may be harmful for older patients with orthostatic hypertension.Circulation2002; 106:196-7.

    [2]Thadani U, Ripley TL. Side effects of using nitrates to treat heart failure and the acute coronary syndromes,unstable angina and acute myocardial infarction. Expert Opin Drug Saf 2007; 6:385-96.

    [3]Piterman L, Zimmet H, Krum H, Tonkin A, Yallop J.Chronic heart failure-optimising care in general practice.Aust Fam Physician2005; 34:547-53.

    [4]Gao XY, Wang HJ, Zhu GQ, Zhang F, Zhang Y, Lu ZH. Effect of implanted chip system on blood pressure regulation in rabbit. J Nanjing Med Univ2004;18:279-82.

    [5]Gao XY, Wang HJ, Zhang Y, Lu ZH, Wang W, Zhu GQ.Resetting blood pressure by a closed-loop implanted chip system in normotensive rats. Life Sci 2006; 78:1129-34

    [6]Uppuluri SC, Storozynsky E, Bisognano JD. Baroreflex device therapy in the treatment of hypertension. Curr Hypertens Rep 2009; 11:69-75.

    [7]Vanoli E, Deferrari GM, Strambabadiale M, Hull SS,Foreman RD, Schwartz PJ. Vagal-stimulation and prevention of sudden-death in conscious dogs with a healed myocardial infarction. Circ Res1991; 68:1471-81.

    [8]Zhang YH, Mowrey KA, Zhuang SW, Wallick DW,Popovic ZB, Mazgalev TN. Optimal ventricular rate slowing during atrial fibrillation by feedback AV nodalselective vagal stimulation. Am J Physiol Heart Circ Physiol 2002; 282:H1102-10.

    [9]Bilgutay AM, Bilgutay IM, Merkel FK, Lillehei CW. Vagal tuning-a new concept in treatment of supraventricular arrhythmias angina pectoris and heart failure. J Thoracic Cardiovasc Surg1968; 56:71-82.

    [10]Buschman HP, Storm CJ, Duncker DJ, Verdouw PD, van der Aa HE, van der Kemp P. Heart rate control via vagus nerve stimulation. Neuromodulation 2006; 9:214-20.

    [11]Tosato M, Yoshida K, Toft E, Nekrasas V, Struijk JJ. Closed-loop control of the heart rate by electrical stimulation of the vagus nerve. Med Biol Eng Comput 2006; 44:161-9.

    [12]Routledge HC, Chowdhary S, Townend JN. Heart rate variability—a therapeutic target? J Clin Pharm Ther2002; 27(2):85-92.

    [13]Schaldach M. What is Closed Loop Stimulation? Progress in Biomedical Research Collection of reprints. May 1996-May 1998; 5:49-55.

    [14]Zhou LM, Zhu GQ, Wang HJ, Zhao CK, Xu Y, Gao XY.Anti-Hypertensive effects of a closed-loop chip system in renovascular hypertensive rats. Med Sci Monit 2008;14:BR153-8.

    [15]MSP430x11x2, MSP430x12x2 Mixed Signal Microcontroller. Texas Instrument Incorporated2004

    [16]MSP430x1xx Family User's Guide. Texas Instrument Incorporated2006

    [17]nRF24L01 Single Chip 2.4GHz Transceiver Product Specification. NORDIC Semiconductor2007

    [18]Duncker DJ, Haitsma DB, Liem DA, Verdouw PD,Merkus D. Exercise unmasks autonomic dysfunction in swine with a recent myocardial infarction. J Cardiovasc Res 2005; 65:889-96.

    各种免费的搞黄视频| 久久香蕉激情| 91精品伊人久久大香线蕉| 别揉我奶头~嗯~啊~动态视频 | 亚洲欧美日韩另类电影网站| 午夜免费鲁丝| 美女扒开内裤让男人捅视频| 亚洲一区中文字幕在线| 亚洲午夜精品一区,二区,三区| 少妇被粗大的猛进出69影院| 法律面前人人平等表现在哪些方面 | 亚洲色图综合在线观看| 亚洲精品在线美女| 色老头精品视频在线观看| 悠悠久久av| 亚洲精品国产区一区二| 亚洲自偷自拍图片 自拍| 亚洲人成77777在线视频| 久久中文字幕一级| 欧美老熟妇乱子伦牲交| 国产精品.久久久| 国产男女超爽视频在线观看| 欧美日韩成人在线一区二区| 国产精品偷伦视频观看了| 国产成人欧美在线观看 | av超薄肉色丝袜交足视频| 精品一区二区三区四区五区乱码| 日本av手机在线免费观看| 手机成人av网站| 韩国精品一区二区三区| www.av在线官网国产| 丝袜人妻中文字幕| 欧美国产精品一级二级三级| 在线 av 中文字幕| 天天躁狠狠躁夜夜躁狠狠躁| 99精国产麻豆久久婷婷| 精品人妻一区二区三区麻豆| 美女高潮到喷水免费观看| 一级片免费观看大全| 日韩一卡2卡3卡4卡2021年| 法律面前人人平等表现在哪些方面 | 成年动漫av网址| 叶爱在线成人免费视频播放| 日本vs欧美在线观看视频| 精品一区二区三区av网在线观看 | 国产日韩欧美视频二区| 他把我摸到了高潮在线观看 | 日韩中文字幕视频在线看片| 精品久久久精品久久久| 国产在线一区二区三区精| 人人妻人人爽人人添夜夜欢视频| 高清视频免费观看一区二区| 在线观看免费日韩欧美大片| 亚洲av电影在线观看一区二区三区| 热99国产精品久久久久久7| 国产男女内射视频| 伦理电影免费视频| 成在线人永久免费视频| 十八禁网站免费在线| 欧美老熟妇乱子伦牲交| 国产日韩欧美视频二区| 在线观看免费午夜福利视频| 高潮久久久久久久久久久不卡| 亚洲欧美色中文字幕在线| 国产日韩一区二区三区精品不卡| 最近最新免费中文字幕在线| 美女中出高潮动态图| av不卡在线播放| 搡老乐熟女国产| 亚洲欧美激情在线| 国产免费av片在线观看野外av| 俄罗斯特黄特色一大片| 亚洲成国产人片在线观看| 精品少妇内射三级| 国产男人的电影天堂91| 少妇被粗大的猛进出69影院| 久久人人爽人人片av| 纵有疾风起免费观看全集完整版| 老司机福利观看| 日韩免费高清中文字幕av| 欧美 日韩 精品 国产| 18禁国产床啪视频网站| 亚洲情色 制服丝袜| 老司机亚洲免费影院| 国产深夜福利视频在线观看| 啦啦啦在线免费观看视频4| 亚洲成人免费电影在线观看| 一边摸一边做爽爽视频免费| 又黄又粗又硬又大视频| 色视频在线一区二区三区| 国产精品99久久99久久久不卡| 亚洲 国产 在线| 99re6热这里在线精品视频| 亚洲精品自拍成人| 性少妇av在线| 国产男人的电影天堂91| 亚洲欧美精品综合一区二区三区| 他把我摸到了高潮在线观看 | 亚洲黑人精品在线| 成人影院久久| 曰老女人黄片| 亚洲精品av麻豆狂野| 国产成人影院久久av| 欧美精品人与动牲交sv欧美| 久久久精品94久久精品| 午夜激情久久久久久久| 成人av一区二区三区在线看 | 免费高清在线观看日韩| 999久久久国产精品视频| 制服人妻中文乱码| 天堂中文最新版在线下载| 亚洲精品国产精品久久久不卡| 99国产精品免费福利视频| 高清视频免费观看一区二区| 老汉色av国产亚洲站长工具| 天天躁夜夜躁狠狠躁躁| 一边摸一边做爽爽视频免费| 少妇粗大呻吟视频| 波多野结衣av一区二区av| 日本精品一区二区三区蜜桃| 高清黄色对白视频在线免费看| 亚洲欧美一区二区三区黑人| 一区二区三区精品91| 国产欧美日韩综合在线一区二区| 女人精品久久久久毛片| 欧美另类一区| 久久久久久久久免费视频了| 久久精品亚洲熟妇少妇任你| 亚洲全国av大片| 一区二区日韩欧美中文字幕| 国产成人一区二区三区免费视频网站| 男人操女人黄网站| 高清av免费在线| 成人18禁高潮啪啪吃奶动态图| 国产在视频线精品| 国产精品一二三区在线看| 国产在线一区二区三区精| 日韩电影二区| 男女床上黄色一级片免费看| 亚洲成人国产一区在线观看| 青春草视频在线免费观看| 大陆偷拍与自拍| 俄罗斯特黄特色一大片| 又大又爽又粗| 日本av免费视频播放| 高清黄色对白视频在线免费看| 国产精品自产拍在线观看55亚洲 | 99久久综合免费| 亚洲专区国产一区二区| 丝袜在线中文字幕| 欧美黑人精品巨大| 国产精品.久久久| 电影成人av| 视频区图区小说| 两个人看的免费小视频| 99久久综合免费| 国产老妇伦熟女老妇高清| 亚洲 国产 在线| 日韩三级视频一区二区三区| 成年人黄色毛片网站| 人人澡人人妻人| 国产欧美日韩一区二区三区在线| 午夜福利,免费看| 亚洲五月婷婷丁香| 国产亚洲午夜精品一区二区久久| 久久 成人 亚洲| 12—13女人毛片做爰片一| 一二三四社区在线视频社区8| 亚洲欧美精品自产自拍| 亚洲国产av新网站| 9191精品国产免费久久| 国产日韩欧美视频二区| avwww免费| 最近最新中文字幕大全免费视频| 国产av又大| 欧美大码av| 中文字幕精品免费在线观看视频| 波多野结衣av一区二区av| 国产成人精品久久二区二区91| e午夜精品久久久久久久| 精品少妇久久久久久888优播| 亚洲五月色婷婷综合| 涩涩av久久男人的天堂| 精品视频人人做人人爽| 亚洲欧美精品自产自拍| 成年人黄色毛片网站| 欧美日韩国产mv在线观看视频| 亚洲精品第二区| 国产成人系列免费观看| 成在线人永久免费视频| 99热网站在线观看| 日日爽夜夜爽网站| 丝袜在线中文字幕| 欧美97在线视频| 免费日韩欧美在线观看| 午夜福利免费观看在线| 亚洲中文av在线| 叶爱在线成人免费视频播放| 精品国产乱码久久久久久男人| 1024视频免费在线观看| 久久精品国产亚洲av高清一级| 成年美女黄网站色视频大全免费| 亚洲精品美女久久av网站| 一区福利在线观看| 欧美日韩中文字幕国产精品一区二区三区 | 美女国产高潮福利片在线看| 亚洲中文日韩欧美视频| 久久精品亚洲熟妇少妇任你| 夜夜骑夜夜射夜夜干| 99九九在线精品视频| 国产成+人综合+亚洲专区| 国产精品久久久久久精品电影小说| 色老头精品视频在线观看| 乱人伦中国视频| 国产精品久久久av美女十八| 日韩欧美国产一区二区入口| 日韩欧美国产一区二区入口| 新久久久久国产一级毛片| av在线老鸭窝| 91成人精品电影| 国产精品国产av在线观看| 成年人免费黄色播放视频| 51午夜福利影视在线观看| 亚洲免费av在线视频| 亚洲成人手机| 又紧又爽又黄一区二区| 99国产极品粉嫩在线观看| 亚洲精品美女久久久久99蜜臀| 国产精品二区激情视频| 精品免费久久久久久久清纯 | 国产免费一区二区三区四区乱码| 亚洲精品在线美女| 在线观看人妻少妇| 亚洲av电影在线进入| 精品熟女少妇八av免费久了| 国产亚洲午夜精品一区二区久久| 国产精品1区2区在线观看. | 一区二区三区激情视频| 少妇精品久久久久久久| 一本一本久久a久久精品综合妖精| 免费看十八禁软件| 久久久久视频综合| 久久综合国产亚洲精品| 国产精品久久久久成人av| 999久久久国产精品视频| 伦理电影免费视频| 午夜免费鲁丝| 久热爱精品视频在线9| 午夜视频精品福利| 啦啦啦免费观看视频1| 午夜免费观看性视频| 深夜精品福利| 欧美xxⅹ黑人| 亚洲免费av在线视频| 无限看片的www在线观看| 动漫黄色视频在线观看| 99国产精品免费福利视频| 啪啪无遮挡十八禁网站| 日韩一卡2卡3卡4卡2021年| 午夜免费鲁丝| 又紧又爽又黄一区二区| 巨乳人妻的诱惑在线观看| h视频一区二区三区| 男女下面插进去视频免费观看| 国产人伦9x9x在线观看| 欧美亚洲日本最大视频资源| 精品一区二区三区四区五区乱码| 老司机深夜福利视频在线观看 | www.熟女人妻精品国产| 狠狠精品人妻久久久久久综合| 成年美女黄网站色视频大全免费| 久久人人爽av亚洲精品天堂| 午夜精品国产一区二区电影| 久久精品亚洲av国产电影网| 欧美亚洲日本最大视频资源| 午夜免费观看性视频| 久久性视频一级片| 精品第一国产精品| 18禁观看日本| 99re6热这里在线精品视频| av福利片在线| 超碰成人久久| 久久av网站| 婷婷丁香在线五月| 人妻人人澡人人爽人人| 亚洲国产成人一精品久久久| 日韩制服丝袜自拍偷拍| 午夜福利乱码中文字幕| 午夜免费观看性视频| 午夜福利影视在线免费观看| 久久av网站| 久久久国产成人免费| 午夜免费成人在线视频| 最近最新中文字幕大全免费视频| 91老司机精品| 天天影视国产精品| 久久久久视频综合| 中文字幕av电影在线播放| 丁香六月天网| 9191精品国产免费久久| 欧美激情久久久久久爽电影 | 亚洲专区中文字幕在线| 精品一品国产午夜福利视频| 久久国产精品影院| 丝袜美腿诱惑在线| 亚洲国产欧美在线一区| 日韩欧美国产一区二区入口| 丝袜美腿诱惑在线| 看免费av毛片| 巨乳人妻的诱惑在线观看| 亚洲人成电影观看| 在线十欧美十亚洲十日本专区| 亚洲男人天堂网一区| 国产精品亚洲av一区麻豆| 久热这里只有精品99| 99精国产麻豆久久婷婷| 国产激情久久老熟女| 一区二区三区激情视频| 欧美精品亚洲一区二区| 丝瓜视频免费看黄片| 欧美激情高清一区二区三区| 国产激情久久老熟女| 欧美日韩亚洲高清精品| 国产精品 国内视频| 50天的宝宝边吃奶边哭怎么回事| av视频免费观看在线观看| 国产av又大| 色婷婷久久久亚洲欧美| 国产男女超爽视频在线观看| 国产成人免费无遮挡视频| 欧美日韩国产mv在线观看视频| 久久毛片免费看一区二区三区| 午夜福利在线免费观看网站| tocl精华| 久久久久久亚洲精品国产蜜桃av| 女人精品久久久久毛片| 一级毛片女人18水好多| 我要看黄色一级片免费的| 99久久99久久久精品蜜桃| 久久国产精品影院| 亚洲专区国产一区二区| 少妇被粗大的猛进出69影院| 狂野欧美激情性xxxx| 手机成人av网站| 国产欧美日韩精品亚洲av| 老司机福利观看| 黑人巨大精品欧美一区二区蜜桃| av超薄肉色丝袜交足视频| 咕卡用的链子| 久久精品aⅴ一区二区三区四区| 国产欧美日韩一区二区精品| 秋霞在线观看毛片| 日韩制服丝袜自拍偷拍| 精品国产乱码久久久久久小说| 日本a在线网址| 精品一区二区三区av网在线观看 | 狂野欧美激情性xxxx| 国产欧美日韩一区二区三区在线| 久久精品国产综合久久久| 免费在线观看完整版高清| 制服人妻中文乱码| 久久久国产欧美日韩av| 最黄视频免费看| 中文精品一卡2卡3卡4更新| 丝瓜视频免费看黄片| 欧美日韩中文字幕国产精品一区二区三区 | 亚洲va日本ⅴa欧美va伊人久久 | 性色av乱码一区二区三区2| 99久久精品国产亚洲精品| 9191精品国产免费久久| 国产深夜福利视频在线观看| 麻豆av在线久日| 嫩草影视91久久| 国产免费福利视频在线观看| 动漫黄色视频在线观看| 中文字幕人妻丝袜制服| 啦啦啦中文免费视频观看日本| 午夜影院在线不卡| bbb黄色大片| 97人妻天天添夜夜摸| 午夜免费观看性视频| 亚洲欧洲日产国产| 男人添女人高潮全过程视频| www.999成人在线观看| 日本av手机在线免费观看| h视频一区二区三区| 久久这里只有精品19| 建设人人有责人人尽责人人享有的| 老司机影院毛片| 午夜免费鲁丝| 午夜福利一区二区在线看| 国产片内射在线| 青春草视频在线免费观看| 日本wwww免费看| 亚洲少妇的诱惑av| 黑丝袜美女国产一区| 少妇猛男粗大的猛烈进出视频| 手机成人av网站| 99热网站在线观看| 999久久久国产精品视频| 99国产精品一区二区蜜桃av | 欧美 亚洲 国产 日韩一| 亚洲精品一二三| 亚洲免费av在线视频| 12—13女人毛片做爰片一| 欧美日韩亚洲综合一区二区三区_| 精品国内亚洲2022精品成人 | 一级a爱视频在线免费观看| 少妇被粗大的猛进出69影院| av在线老鸭窝| 久久久久精品人妻al黑| 精品国产一区二区三区久久久樱花| 亚洲性夜色夜夜综合| 精品第一国产精品| 精品一品国产午夜福利视频| 大片电影免费在线观看免费| 精品高清国产在线一区| 老司机影院毛片| 啦啦啦在线免费观看视频4| 久久av网站| 久久久精品国产亚洲av高清涩受| 国产成人精品久久二区二区91| 国产av精品麻豆| 久久性视频一级片| 成人影院久久| svipshipincom国产片| 精品久久久精品久久久| 国产深夜福利视频在线观看| 久久99一区二区三区| 色播在线永久视频| 啦啦啦 在线观看视频| 亚洲欧洲日产国产| 91成人精品电影| 男女边摸边吃奶| 女性被躁到高潮视频| 国产av精品麻豆| 男男h啪啪无遮挡| 欧美变态另类bdsm刘玥| 成人国语在线视频| 亚洲少妇的诱惑av| 久久精品亚洲av国产电影网| 国产无遮挡羞羞视频在线观看| 成年人黄色毛片网站| 国产亚洲av高清不卡| 婷婷色av中文字幕| 少妇被粗大的猛进出69影院| bbb黄色大片| 亚洲国产欧美日韩在线播放| 日本a在线网址| 日韩免费高清中文字幕av| 搡老岳熟女国产| 国产精品亚洲av一区麻豆| 一边摸一边抽搐一进一出视频| 黄频高清免费视频| 欧美乱码精品一区二区三区| 亚洲九九香蕉| 国产精品一区二区精品视频观看| 免费观看av网站的网址| 亚洲va日本ⅴa欧美va伊人久久 | 国产av一区二区精品久久| 黄色片一级片一级黄色片| 国产成人欧美| 两性夫妻黄色片| 久久人人爽人人片av| 中亚洲国语对白在线视频| 不卡av一区二区三区| 欧美日韩亚洲综合一区二区三区_| 午夜91福利影院| 色94色欧美一区二区| 免费日韩欧美在线观看| 国产免费现黄频在线看| 美女国产高潮福利片在线看| 香蕉国产在线看| 自拍欧美九色日韩亚洲蝌蚪91| 亚洲伊人色综图| 人人妻人人澡人人看| 一级a爱视频在线免费观看| 一区在线观看完整版| 女人爽到高潮嗷嗷叫在线视频| 国产精品.久久久| 免费一级毛片在线播放高清视频 | 黄片小视频在线播放| 精品第一国产精品| 欧美人与性动交α欧美精品济南到| 激情视频va一区二区三区| 最近最新免费中文字幕在线| 悠悠久久av| 一区二区日韩欧美中文字幕| 午夜成年电影在线免费观看| 欧美黑人精品巨大| 桃花免费在线播放| 日韩欧美一区视频在线观看| 精品国产一区二区久久| 老熟女久久久| 大片电影免费在线观看免费| 国产淫语在线视频| 国产亚洲午夜精品一区二区久久| 久久久国产欧美日韩av| 首页视频小说图片口味搜索| 亚洲欧美一区二区三区久久| 男女无遮挡免费网站观看| 桃红色精品国产亚洲av| 在线十欧美十亚洲十日本专区| 亚洲精品美女久久av网站| 国产av又大| 欧美精品亚洲一区二区| 成人亚洲精品一区在线观看| 熟女少妇亚洲综合色aaa.| 欧美97在线视频| 最近最新免费中文字幕在线| 午夜福利影视在线免费观看| 久久毛片免费看一区二区三区| 一进一出抽搐动态| 免费日韩欧美在线观看| 国产一区二区三区综合在线观看| 中文字幕另类日韩欧美亚洲嫩草| 正在播放国产对白刺激| 我要看黄色一级片免费的| 亚洲精品第二区| 国产黄色免费在线视频| 狠狠精品人妻久久久久久综合| 久久精品亚洲av国产电影网| 12—13女人毛片做爰片一| 在线观看免费午夜福利视频| 午夜视频精品福利| 一级片'在线观看视频| www.自偷自拍.com| 国产成人免费观看mmmm| 久久久国产欧美日韩av| 国产国语露脸激情在线看| 国产亚洲av片在线观看秒播厂| av超薄肉色丝袜交足视频| 又紧又爽又黄一区二区| 亚洲熟女精品中文字幕| 成年人午夜在线观看视频| 丝袜美腿诱惑在线| 成人黄色视频免费在线看| 999久久久国产精品视频| 少妇人妻久久综合中文| 午夜精品国产一区二区电影| 日本猛色少妇xxxxx猛交久久| 男女无遮挡免费网站观看| 国产精品欧美亚洲77777| 国产主播在线观看一区二区| 国产高清videossex| 亚洲精品美女久久久久99蜜臀| 桃红色精品国产亚洲av| 飞空精品影院首页| 国产一区二区在线观看av| 真人做人爱边吃奶动态| 亚洲性夜色夜夜综合| 欧美国产精品va在线观看不卡| 俄罗斯特黄特色一大片| 动漫黄色视频在线观看| 深夜精品福利| 丁香六月欧美| 香蕉丝袜av| 高潮久久久久久久久久久不卡| 亚洲av日韩在线播放| 又黄又粗又硬又大视频| 久久免费观看电影| 欧美久久黑人一区二区| 人妻 亚洲 视频| netflix在线观看网站| 欧美在线黄色| 一级毛片电影观看| 一区二区三区乱码不卡18| 国产成人精品无人区| 水蜜桃什么品种好| 丝袜人妻中文字幕| 欧美 日韩 精品 国产| 欧美日韩中文字幕国产精品一区二区三区 | 国产一区二区三区综合在线观看| 伦理电影免费视频| 乱人伦中国视频| 国产成人啪精品午夜网站| 老司机亚洲免费影院| 精品国产一区二区三区久久久樱花| 欧美日韩成人在线一区二区| 欧美日韩亚洲国产一区二区在线观看 | 久久久久视频综合| 中文字幕制服av| 一区在线观看完整版| 99热全是精品| 国产成人精品无人区| 少妇 在线观看| cao死你这个sao货| 亚洲精品成人av观看孕妇| av视频免费观看在线观看| av片东京热男人的天堂| 国产成人av教育| 一本—道久久a久久精品蜜桃钙片| 亚洲精品国产av蜜桃| 欧美日韩视频精品一区| 日韩 亚洲 欧美在线| 久久99一区二区三区| 老熟女久久久| 伊人亚洲综合成人网| 精品人妻一区二区三区麻豆| 黄片大片在线免费观看| 97精品久久久久久久久久精品| 久久性视频一级片| 老熟妇乱子伦视频在线观看 | 午夜免费成人在线视频| 狂野欧美激情性xxxx| 国产野战对白在线观看| 久久久久精品国产欧美久久久 | 侵犯人妻中文字幕一二三四区| 丝袜美腿诱惑在线| 亚洲国产成人一精品久久久| 少妇的丰满在线观看| 涩涩av久久男人的天堂| 2018国产大陆天天弄谢| 免费黄频网站在线观看国产| 精品一区在线观看国产|